wiki:Docs/User/FullUpdate/VersionHistory/OptionCards

Option Card Update History

825 Option cards cannot be updated automatically using the USB update pack. Except for the DLC card, firmware updates to the option cards require an ISP (In system programming) hardware interface and appropriate software to perform the update.

Click this link for information about updating the DLC option card

DAC (Digital to Analog)

Ver: 0.01.000 February 17, 2011

Original Release

DIO (Digital I/O)

Ver: 0.05.000 November 8, 2011

Added pulse/PWM output feature.

Ver: 0.04 May 13, 2010

Added counting feature.

Ver: 0.03 November 18, 2008

Initial release

DLC (Digital Load Cell)

Ver: 1.00.010 May 1, 2019

Improved CAN bus recovery in case of interruption of V+ or V- line. Added support for new diagnostics screen "DLC Card Communications Error Counts". Fixed issue that iSite network connection may not recover automatically if network is interrupted and then reconnected, or if DHCP assigns a new IP address.

Ver: 1.00.001 October 29, 2018

Fixed issue that if previously configured for more cells, and then changed to fewer cells the calibration may not properly set zero.

Ver: 1.00.000 October 25, 2018

Initial Release

ICAN (ICAN Board)

Ver: 1.00 May 27, 2010

Original Release

SIB (Scale Interface Board)

Ver: 1.10.000 November 17, 2023

Fix for previous version 1.09. Auto-zero could sometimes cause delay of several seconds writing EEPROM and in some cases especially if power cycled memory could be corrupted causing loss of calibration and all settings to return to defaults.

Ver: 1.09.000 May 16, 2022

No operational change, just recompiled and linked for DN microcontroller. It is still compatible with DZ microcontroller.

Ver: 1.08.000 June 29, 2016

Improved zeroing to not write updates to the internal EEPROM as often. Previously, in some cases with auto-zero enabled frequent zeroing could wear out the EEPROM prematurely.

Ver: 1.07.000 July 3, 2012

Changed version number to be consistent with new standard. Operation is identical to version 0.07 except for output of version information.

Ver: 0.07 May 6, 2011

Added FIR filter option. Break Range 0 = always on, Value 2 - 64.

Ver: 0.06 November 18, 2008

Clear weight average after calibration

Ver: 0.05 September 17, 2008

Changed default configuration parameters for board test.

Ver: 0.04 July 31, 2008

Added units labels to board parameters.

Last modified 9 months ago Last modified on 02/05/24 13:06:53
Note: See TracWiki for help on using the wiki.